WEKO3
アイテム
A Mobile Cough Strength Evaluation Device Using Cough Sounds
https://hcu.repo.nii.ac.jp/records/190
https://hcu.repo.nii.ac.jp/records/190a3f3e793-6df6-4d7b-8630-77b8bd71acb7
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2023-03-14 | |||||
タイトル | ||||||
タイトル | A Mobile Cough Strength Evaluation Device Using Cough Sounds | |||||
言語 | en | |||||
タイトル | ||||||
タイトル | A Mobile Cough Strength Evaluation Device Using Cough Sounds | |||||
言語 | en | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | cough strength | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | cough sound | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | cough peak flow | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | mobile device | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | iOS | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | open access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_abf2 | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Although cough peak flow (CPF) is an important measurement for evaluating the risk of cough dysfunction, some patients cannot use conventional measurement instruments, such as spirometers, because of the configurational burden of the instruments. Therefore, we previously developed a cough strength estimation method using cough sounds based on a simple acoustic and aerodynamic model. However, the previous model did not consider age or have a user interface for practical application. This study clarifies the cough strength prediction accuracy using an improved model in young and elderly participants. Additionally, a user interface for mobile devices was developed to record cough sounds and estimate cough strength using the proposed method. We then performed experiments on 33 young participants (21.3 ± 0.4 years) and 25 elderly participants (80.4±6.1 years) to test the effect of age on the CPF estimation accuracy. The percentage error between the measured and estimated CPFs was approximately 6.19%. In addition, among the elderly participants, the current model improved the estimation accuracy of the previous model by a percentage error of approximately 6.5% (p < 0.001). Furthermore, Bland-Altman analysis demonstrated no systematic error between the measured and estimated CPFs. These results suggest that the developed device can be applied for daily CPF measurements in clinical practice. | |||||
書誌情報 |
Sensors 巻 18, 号 Issue 11, p. 3810(1)-3810(13), 発行日 2018-11-07 |
|||||
出版者 | ||||||
出版者 | MDPI AG | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 14248220 | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | info:doi/10.3390/s18113810 | |||||
権利 | ||||||
権利情報 | (C)MDPI AG | |||||
フォーマット | ||||||
内容記述タイプ | Other | |||||
内容記述 | application/pdf | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
日本十進分類法 | ||||||
主題Scheme | NDC | |||||
主題 | 493.3|007.64 |